NEW
Specification
Next-gen electrosurgery: Independent endoscope port, auto-mode switching, precise power control, error monitoring, and waterproof panel. Optimal safety, efficiency, and versatility in a single unit
Scope Application of MCS0431 Electrosurgery Unit
Electrosurgery Unit-1
Electrosurgical Unit is suitable for various surgeries on cutting or coagulation, including general surgery, heart, gynecology, anorectal surgery, orthopedics,thoracic surgery, tumor, etc.
Electrosurgical Unit also used with surgeries of endoscope, hysteroscope, laparoscope ENT endoscope, etc. Bipolar can be used in the fine surgeries of microsurgery, neurology, ENT and ophthalmology, hand surgery, etc.
Electrosurgery Unit Advantages
1. Independent output port of endoscope and intelligent conversion key, which can enter endoscopic mode automatically after start-up.
2. Microcomputer control system, power-off protection, and memory function of keeping last usage data after a restart.
3. Automatic adjustment on output power. It can be adjusted automatically according to the change in the density of the tissue. Stable output power to keep the minimum wastage
4. Automatic conversion on monopolar and bipolar
5. Automatic monitoring of on-off, automatic detection, and error prompts in work.
6. The monitoring circuit system on contact quality of neutral electrodes can test and evaluate the safety and effectiveness of the contact area between the electrode plate and skin.
7. The system can cut off output automatically and give an alarm, in case of finds that the contact area is at a dangerous level. It can test and use a negative plate of monopolar or bipolar effectively.
8. Operating panel with keys, high definition, and large digital. It has different sound and light indicators during operation. Waterproof operating keys are easy to clean and maintain.
9. Independent three-way power output. This increases the convenience and security of operations.
10. It can be operated and performed well underwater even when dissecting and removal of fat tissue.
11. Fully suspended power output. There are two independent and isolated application sections to prevent defibrillation interference (monopolar and bipolar).
Technical Details
Electrosurgery Unit has 5 Working Modes
Mono Cut Cutting: 400W Blend: 150W
MonoCoag Soft Coag: 100W Strong Coag: 80W
It will beep Bipolar Coag: 50W
Technical Parameters of Electrosurgery Unit
Ambient temperature range 10℃~40℃
Relative humidity range 30%~75%
Atmospheric pressure range 700hpa~1060hpa
Power supply 220V/110V,50Hz
Working frequency 360kHz~460kHz
Type of equipment CF
The power consumption of the whole device is less than 1000VA. (Cutting function:400W)